This handheld device is specifically meant to calibrate process measuring equipment.
Thermocouples, RTD's, voltage, current and ohms can be measured and sourced with the touch
of a key. At the heart of the system is a 24 bit Sigma Delta ADC of 0.001% accuracy with auto zero
facility and calibration against built in reference standards.
Applications
-Laboratories & RnD units;
-Manufacturers of all process control equipment and industries
where sensor o/p calibration is required on a periodic basis
to satisfy quality norms
Detailed Specifications
| |
Range |
Resolution |
Accuracy |
| Thermocouples |
E,J,K,T,R,S,B,N |
0.1 C |
Max errors:
E,J,K,T: 0.2 C
R,S,B: 1.2 C
N: 0.4 C |
| RTDs |
PT 385 (100Ω, 200Ω, 500Ω, 1000Ω)
PT 392 (100Ω, 200Ω, 500Ω, 1000Ω) |
0.01 C |
0.05 C |
| mV |
0 to 120 mV |
1µV |
0.01%
+ 10µV |
| Volt |
0 to 1V & 0 to 5V |
10µV & 100µV |
0.015%
+ 200µV &
0.015% + 1mV
|
| mA current |
0 to 24 mA |
1µA |
+
0.01% |
| Ohms |
0 to 400Ω & 0 to 4KΩ |
0.01Ω & 0.1Ω |
0.025% |
|
|
|
|
|
|